A handheld polarimetric imaging device and calibration technique for accurate mapping of terahertz Stokes vectors.
Ontology highlight
ABSTRACT: In recent years, handheld and portable terahertz instruments have been in rapid development for various applications ranging from non-destructive testing to biomedical imaging and sensing. For instance, we have deployed our Portable Handheld Spectral Reflection (PHASR) Scanners for in vivo full-spectroscopic imaging of skin burns in large animal models in operating room settings. In this paper, we debut the polarimetric version of the PHASR Scanner, and describe a generalized calibration technique to map the spatial and spectral dependence of the Jones matrix of an imaging scanner across its field of view. Our design is based on placement of two orthogonal photoconductive antenna (PCA) detectors separated by a polarizing beam splitter in the PHASR Scanner housing. We show that as few as th
SUBMITTER: Harris ZB
PROVIDER: S-EPMC11292021 | biostudies-literature | 2024 Jul
REPOSITORIES: biostudies-literature