Accuracy of intraocular lens calculation formulas based on swept-source OCT biometer in cataract patients with phakic intraocular lens.


ABSTRACT:

Purpose

To research the accuracy of intraocular lens (IOL) calculation formulas and investigate the effect of anterior chamber depth (ACD) and lens thickness (LT) measured by swept-source optical coherence tomography biometer (IOLMaster 700) in patients with posterior chamber phakic IOL (PC-pIOL).

Methods

Retrospective case series. The IOLMaster 700 biometer was used to measure axial length (AL) and anterior segment parameters. The traditional formulas (SRK/T, Holladay 1 and Haigis) with or without Wang-Koch (WK) AL adjustment, and new-generation formulas (Barret Universal II [BUII], Emmetropia Verifying Optical [EVO] v2.0, Kane, Pearl-DGS) were utilized in IOL power calculation.

Results

This study enrolled 24 eyes of 24 patients undergoing combined PC-pIOL removal an