Standardization of body composition parameters between GE Lunar iDXA and Hologic Horizon A and their clinical impact.


ABSTRACT: Body composition (BC) measured by DXA differs between devices. We aimed to compare regional and total BC measurements assessed by the Hologic Horizon A and the GE Lunar iDXA devices; to determine device-specific calibration equations for each BC parameter; and to assess the impact of this standardization procedure on the assessment of sarcopenia, lipedema, obesity, and cardiovascular risk with DXA. A total of 926 postmenopausal women (aged 72.9 ± 6.9 yr, height 160.3 ± 6.6 cm, weight 66.1 ± 12.7 kg) underwent BC assessment on each device within 1 h, following the ISCD guidelines. The included sample was split into 80% train and 20% test datasets stratified by age, height, and weight. Inter-device differences in BC parameters were assessed with Bland-Altman analysis, Pearson or Spearman cor