Use of Nonhuman Sera as a Highly Cost-Effective Internal Standard for Quantitation of Multiple Human Proteins Using Species-Specific Tryptic Peptides: Applicability in Clinical LC-MS Analyses.


ABSTRACT: Quantitation of proteins using liquid chromatography-tandem mass spectrometry (LC-MS/MS) is complex, with a multiplicity of options ranging from label-free techniques to chemically and metabolically labeling proteins. Increasingly, for clinically relevant analyses, stable isotope-labeled (SIL) internal standards (ISs) represent the "gold standard" for quantitation due to their similar physiochemical properties to the analyte, wide availability, and ability to multiplex to several peptides. However, the purchase of SIL-ISs is a resource-intensive step in terms of cost and time, particularly for screening putative biomarker panels of hundreds of proteins. We demonstrate an alternative strategy utilizing nonhuman sera as the IS for quantitation of multiple human proteins. We demonstrate the e

SUBMITTER: Williams G 

PROVIDER: S-EPMC11301776 | biostudies-literature | 2024 Aug