Unknown

Dataset Information

Serial Vascular Responses of Balloon-Expandable Stent With Biodegradable Film-Type Graft in a Rabbit Iliac Artery Dissection Model (BioGard Study).


ABSTRACT:

Background and objectives

Arterial dissection during endovascular therapy rarely occurs but can be lethal. A fabric-based covered graft stents yield poor clinical outcomes. A novel balloon-expandable stent with biodegradable film graft for overcoming these issues was evaluated in a rabbit iliac artery model.

Method

Eighteen rabbits with iliac artery dissections were induced by balloon over-inflation on angiography (Ellis type 2 or 3) and treated using the test device (3.0×24 mm). Subsequently, survived twelve animals underwent histologic examinations and micro-computed tomography (CT) at 0, 2, 4, and 8 weeks and 3, 6, 9, and 12 months and angiography at one-year.

Results

There were no adverse cardiovascular events during the one-year. Early-stage histologic examinatio

SUBMITTER: Park SM 

PROVIDER: S-EPMC11306422 | biostudies-literature | 2024 Aug

REPOSITORIES: biostudies-literature