High-Throughput Raman Spectroscopy by Horizontally Shifted Collection Fibers.


ABSTRACT: Optical fiber probe-based Raman spectroscopy systems are widely used for in situ measurements ranging from material characterization to biomedical applications. However, small Raman cross sections necessitate the use of high-power lasers or long exposure times that limit Raman's larger application to multiple research fields. This limitation can be overcome by collecting more Raman photons through additional collection fibers with taller detectors. This system configuration requires replacement of the detector and modification of the spectrograph to incorporate larger optical components, making it a costly and cumbersome option. In probe-based Raman systems, a typical detector image shows stacked collection fibers on the vertical axis and Raman spectra on the horizontal axis. While the ver