Stretched to the Limit: Comparing Polytetrafluoroethylene-Covered Endovascular Stents Through Serial Dilations.


ABSTRACT:

Background

Covered stents are used during congenital cardiac interventions to treat stenotic or injured vessels or to exclude unwanted vascular connections. The ability to postdilate a stented vessel to keep pace with somatic growth is critical in children. In this study, we aimed to compare in vitro performance of 2 brands of covered stents during serial dilations to demonstrate the threshold for stent fracture and polytetrafluoroethylene tear and define recoil and foreshortening characteristics.

Methods

iCast and VBX stents of various sizes were measured before and after expansion and through serial dilations. Dilations were performed at 2-mm increments until stent fracture, polytetrafluoroethylene tear, and "napkin-ring" formation, to a maximum of 22-mm diameter.