Comparison of Glaucoma Detection Performance of Binocular Perimetry Screening Program Using imo Perimetry With Frequency Doubling Technology.


ABSTRACT:

Purpose

We aimed to preliminarily compare the glaucoma detection accuracy of a head-mounted binocular visual perimeter "imo" screening program (ISP) with that of frequency doubling technology (FDT).

Methods

This multicenter, diagnostic accuracy study based on prospectively collected data included 76 non-glaucoma (including pre-perimetric glaucoma) eyes and 92 glaucomatous eyes from patients visiting two hospitals. Patients underwent ISP and FDT (C-20-1 screening program) on the same day. Diagnostic efficacy was evaluated using receiver operating characteristic curves and areas under the curve (AUCs). In addition, we compared the ISP and FDT testing times.

Results

AUC values for ISP versus FDT were as follows: (1) mild-stage glaucoma (mean deviation [MD] > -6 dB), 0.82