Compressed SENSitivity Encoding (SENSE): Qualitative and Quantitative Analysis.


ABSTRACT:

Background

This study aimed to qualitatively and quantitatively evaluate T1-TSE, T2-TSE and 3D FLAIR sequences obtained with and without Compressed-SENSE technique by assessing the contrast (C), the contrast-to-noise ratio (CNR) and the signal-to-noise ratio (SNR).

Methods

A total of 142 MRI images were acquired: 69 with Compressed-SENSE and 73 without Compressed-SENSE. All the MRI images were contoured, spatially aligned and co-registered using 3D Slicer Software. Two radiologists manually drew 12 regions of interests on three different structures of CNS: white matter (WM), grey matter (GM) and cerebrospinal fluid (CSF).

Results

C values were significantly higher in Compressed-SENSE T1-TSE compared to No Compressed-SENSE T1-TSE for three different structures of the C

SUBMITTER: Picchi E 

PROVIDER: S-EPMC11311492 | biostudies-literature | 2024 Aug

REPOSITORIES: biostudies-literature