Between-Session Reliability of Field-Based Assessments of Isometric Neck Strength.
Ontology highlight
ABSTRACT: The reliability of the fixed-frame dynamometer for measuring isometric neck strength is established, but with limited field-based applications. This study aimed to establish the inter- and intra-session reliability of the peak force for neck flexors, extensors, and side flexors using the VALD ForceFrame and DynaMo and the force-time characteristics in the quadruped position (ForceFrame). Twenty-seven recreationally active males performed three repetitions of isometric neck flexion, extension, and side flexion over two sessions in random order using the VALD ForceFrame and DynaMo. Both devices demonstrated acceptable reliability, with the Forceframe ICC > 0.8 and CV% < 13.8% and the DynaMo ICC > 0.76 and CV% < 13.8%. No systematic or proportional differences were found using the Passing-Bab