A computational study of artery curvature and endograft oversize influence on seal zone behavior in endovascular aortic repair.
Ontology highlight
ABSTRACT: Thoracic endovascular aortic repair (TEVAR) is a minimally invasive procedure involving the placement of an endograft inside the dissection or an aneurysm to direct blood flow and prevent rupture. A significant challenge in endovascular surgery is the geometrical mismatch between the endograft and the artery, which can lead to endoleak formation, a condition where blood leaks between the endograft and the vessel wall. This study uses computational modeling to investigate the effects of artery curvature and endograft oversizing, the selection of an endograft with a larger diameter than the artery, on endoleak creation. Finite element analysis is employed to simulate the deployment of endografts in arteries with varying curvature and diameter. Numerical simulations are conducted to assess th
SUBMITTER: Donik Z
PROVIDER: S-EPMC11317088 | biostudies-literature | 2024 Aug
REPOSITORIES: biostudies-literature
ACCESS DATA