Clinical Validation of Carotid-Femoral Pulse Wave Velocity Measurement Using a Multi-Beam Laser Vibrometer: The CARDIS Study.


ABSTRACT:

Background

Carotid-femoral pulse wave velocity (cfPWV) is the gold standard for noninvasive arterial stiffness assessment, an independent predictor of cardiovascular disease, and a potential parameter to guide therapy. However, cfPWV is not routinely measured in clinical practice due to the unavailability of a low-cost, operator-friendly, and independent device. The current study validated a novel laser Doppler vibrometry (LDV)-based measurement of cfPWV against the reference technique.

Methods

In 100 (50 men) hypertensive patients, cfPWV was measured using applanation tonometry (Sphygmocor) and the novel LDV device. This device has 2 handpieces with 6 laser beams each that simultaneously measure vibrations from the skin surface at carotid and femoral sites. Pulse wave veloc