Improved efficiency using sequential automated immunoassays for syphilis screening in blood donors.
Ontology highlight
ABSTRACT: Using sequential immunoassays for the screening of blood donors is well described for viral serology testing but not for the screening of syphilis. In this study, we report the evaluation results and 2-year sequential testing data using two highly sensitive automated serology assays, the Alinity s Syphilis chemiluminescent immunoassay for screening, with all repeatedly reactive samples then tested on the Elecsys Syphilis electrochemiluminescence immunoassay. We screened 1,767,782 blood donor samples between 7 July 2021 and 6 July 2023 and found the Alinity false-positive rate to be low at 0.08% (1,456/1,767,782). The common false-positive rate between the two assays was also low (3.83%, 58/1,514). Concordantly reactive samples were further tested using a Treponema pallidum particle
SUBMITTER: Cheng A
PROVIDER: S-EPMC11323462 | biostudies-literature | 2024 Aug
REPOSITORIES: biostudies-literature
ACCESS DATA