Flash-kinetics as a complementary analytical tool in PAM fluorimetry.


ABSTRACT: A new measuring system based on the already existing Multi-Color-PAM Fluorimeter (Schreiber et al. in Photosynth Res 113:127-144, 2012) was developed that in addition to standard PAM measurements enables pump-and-probe flash measurements and allows simultaneous measurements of the changes in chlorophyll fluorescence yield (F) during application of saturating flashes (ST). A high-power Chip-on-Board LED array provides ST flashes with close to rectangular profiles at wide ranges of widths (0.5 µs to 5 ms), intensities (1.3 mmol to 1.3 mol 440 nm quanta m-2 s-1) and highly flexible repetition times. Using a dedicated rising-edge profile correction, sub-µs time resolution is obtained for assessment of initial fluorescence and rise kinetics. At maximal to moderate flash in

SUBMITTER: Klughammer C 

PROVIDER: S-EPMC11324780 | biostudies-literature | 2024 Sep