Unknown

Dataset Information

Control of mean arterial pressure using a closed-loop system for norepinephrine infusion in severe brain injury patients: the COMAT randomized controlled trial.


ABSTRACT: Brain injury patients require precise blood pressure (BP) management to maintain cerebral perfusion pressure (CPP) and avoid intracranial hypertension. Nurses have many tasks and norepinephrine titration has been shown to be suboptimal. This can lead to limited BP control in patients that are in critical need of cerebral perfusion optimization. We have designed a closed-loop vasopressor (CLV) system capable of maintaining mean arterial pressure (MAP) in a narrow range and we aimed to assess its performance when treating severe brain injury patients. Within the first 48 h of intensive care unit (ICU) admission, 18 patients with a severe brain injury underwent either CLV or manual norepinephrine titration. In both groups, the objective was to maintain MAP in target (within ± 5 mmHg of a pred

SUBMITTER: Joosten A 

PROVIDER: S-EPMC11330589 | biostudies-literature | 2024 Feb

REPOSITORIES: biostudies-literature