Evaluation of in vitro irradiation setup: Designed for the horizontal beamline at the Danish Centre for Particle Therapy.


ABSTRACT:

Background

Radiobiological experimental setups are challenged by precise sample positioning along depth dose profile, scattering conditions, and practical difficulties that must be addressed in individual designs. The aim of this study was to produce cell survival curves with several irradiation modalities, by using a setup designed at the Danish Centre for Particle Therapy (DCPT) for in vitro proton irradiations using a horizontal beam line and thereby evaluating the setups use for in vitro irradiations experiments.

Materials and methods

The setup is a water phantom suitable for in vitro research with multiple irradiation modalities, in particular the pencil scanning proton beam available from a horizontal experimental beamline. The phantom included a water tank of 39.0 × 1