Shear-Wave Elastography of the Achilles tendon: reliability analysis and impact of parameters modulating elasticity values.


ABSTRACT:

Purpose

Shear wave elastography (SWE) has seen many advancements in Achilles tendon evaluation in recent years, yet standardization of this technique is still problematic due to the lack of knowledge regarding the optimal way to perform the examination. The purpose of this study was to evaluate the effects of ankle position, probe frequency and physical effort on the shear modulus of the Achilles tendon, but also to determine the intra and inter-observer reliability of the technique.

Methods

37 healthy volunteers were included; SWE protocol was performed by two examiners. We analyzed the shear modulus of the tendon with the ankle in neutral, maximum dorsiflexion and maximum plantar flexion using two different high frequency probes. Afterwards, the subjects performed a brief