Primary stability evaluation of different morse cone implants in low-density artificial bone blocks: A comparison between high-and low-speed drilling.


ABSTRACT: This study aimed to evaluate various biomechanical parameters associated with the primary stability of Maestro and Due Cone implants placed in low-density artificial bones, prepared using high-speed drilling with irrigation and low-speed drilling without irrigation. The insertion torque (IT), removal torque (RT), and implant stability quotient (ISQ) values were recorded for Maestro and Due Cone implants placed in low-density polyurethane blocks (10 and 20 pounds per cubic foot (PCF) with and without a cortical layer) prepared using high-speed and low-speed with or without irrigation using a saline solution, respectively. A three-way ANOVA model and Tukey's post-hoc test were conducted, presenting data as means and standard deviations. P-values equal to or less than 0.05 were conside