Unknown

Dataset Information

Modes of administration of nitric oxide devices and ventilators flow-by impact the delivery of pre-determined concentrations.


ABSTRACT:

Background

Nitric oxide (NO) is a strong vasodilator, selectively directed on pulmonary circulation through inhaled administration. In adult intensive care units (ICU), it is mainly used for refractory hypoxemia in mechanically ventilated patients. Several medical delivery devices have been developed to deliver inhaled nitric oxide (iNO). The main purpose of those devices is to guarantee an accurate inspiratory NO concentration, whatever the ventilator used, with NO2 concentrations lower than 0.3 ppm. We hypothesized that the performances of the different available iNO delivery systems could depend on their working principle and could be influenced by the ventilator settings. The objective of this study was to assess the accuracy of seven different iNO-devices combined w

SUBMITTER: Vuillermoz A 

PROVIDER: S-EPMC11339004 | biostudies-literature | 2024 Aug

REPOSITORIES: biostudies-literature