ABSTRACT: Purpose
We aim to compare the imaging performance of a cone-beam CT (CBCT) imaging system with noncircular scan protocols (sine-on-sphere) to a conventional circular orbit.Approach
A biplane C-arm system (ARTIS Icono; Siemens Healthineers) capable of circular and noncircular CBCT acquisition was used, with the latter orbit (sine-on-sphere, "Sine Spin") executing a sinusoidal motion with ±10 deg tilt amplitude over the half-scan orbit. A test phantom was used for the characterization of image uniformity, noise, noise-power spectrum (NPS), spatial resolution [modulation transfer function (MTF) in axial and oblique directions], and cone-beam artifacts. Findings were interpreted using an anthropomorphic head phantom with respect to pertinent tasks in skull base neurosurgery.