Unknown

Dataset Information

Microfluidic QCM enables ultrahigh Q-factor: a new paradigm for in-liquid gravimetric sensing.


ABSTRACT: Acoustic gravimetric biosensors attract attention due to their simplicity, robustness, and low cost. However, a prevailing challenge in these sensors is dissipation which manifests in a low quality factor (Q-factor), which limits their sensitivity and accuracy. To mitigate dissipation of acoustic sensors in liquid environments we introduce an innovative approach in which we combine microfluidic channels with gravimetric sensors. To implement this novel paradigm we chose the quartz crystal microbalance (QCM) as our model system, owing to its wide applicability in biosensing and the relevance of its operating principles to other types of acoustic sensors. We postulate that the crucial determinant for enhancing performance lies in the ratio between the width of the microfluidic channels and t

SUBMITTER: Zhao Y 

PROVIDER: S-EPMC11347674 | biostudies-literature | 2024 Aug

REPOSITORIES: biostudies-literature