ABSTRACT: Background
Cerclage wiring is commonly used for treating fractures; however, it has several limitations, including mechanical weakness, decreased blood circulation, and technical complexity. In this study, we developed an implant using a shape memory alloy (SMA) and tested its efficacy in treating Vancouver type B1 (VB1) periprosthetic femoral fractures (PFFs) in a canine model.Methods
The mid-diaphyseal fracture models underwent reduction via the SMA plate (SMA group) or the cerclage cable plate (cable group) method in randomly selected pelvic limbs. An intraoperative evaluation was conducted to assess the surgical time and difficulty related to implant fitting. Clinical assessments, radiography, microcomputed tomography (micro-CT), histological analysis, positron emission