ABSTRACT: Background
High-flow nasal cannula (HFNC) systems employ different methods to provide aerosol to patients. This study compared delivery efficiency, particle size, and regional deposition of aerosolized bronchodilators during HFNC in neonatal, pediatric, and adult upper-airway and lung models between a proximal aerosol adapter and distal aerosol circuit chamber.Methods
A filter was connected to the upper airway to a spontaneously breathing lung model. Albuterol was nebulized using the aerosol adapter and circuit at different clinical flow settings. The aerosol mass deposited in the upper airway and lung was quantified. Particle size was measured with a laser diffractometer. Regional deposition was assessed with a gamma camera at each nebulizer location and patient model with