Performance Evaluation of Open Channel Buhlmann Fecal Calprotectin Turbo Assay on Abbott Alinity C Analyzer.
Ontology highlight
ABSTRACT: Inflammatory bowel disease (IBD) is characterized by chronic inflammation of the gastrointestinal (GI) tract. Fecal calprotectin (fCAL) is a noninvasive laboratory test used in the diagnosis and monitoring of IBDs such as Crohn's disease and ulcerative colitis. The fCAL send-out test that our facility has been offering so far uses an ELISA-based method. In the current study, we sought to validate the performance of a Buhlmann fCAL turbo assay in an automated Abbott Alinity C analyzer (AFCAL) in our core laboratory. Five-day imprecision studies showed good performance for both within-run (5.3%) and between-day (2.5%) measurements. The reportable range was verified as 30-20,000 µg/g. Deming regression and Bland-Altman analysis indicated a strong correlation of r = 0.99 with a low, acceptable
SUBMITTER: Sataranatarajan K
PROVIDER: S-EPMC11353904 | biostudies-literature | 2024 Aug
REPOSITORIES: biostudies-literature
ACCESS DATA