ABSTRACT: Background
To treat stenosed coronary arteries, percutaneous transluminal coronary angioplasty (PTCA) balloon catheters must combine pushability, trackability, crossability, and rewrap behavior. The existing anatomic track model (ASTM F2394) for catheter testing lacks 3D morphology, vessel tortuosity, and compliance, making evaluating performance characteristics difficult. This study aimed to develop a three-dimensional patient-specific phantom (3DPSP) for device testing and safe training for interventional cardiologists.Methods
A range of silicone materials with different shore hardnesses (00-30-45 A) and wall thicknesses (0.5 mm, 1 mm, 2 mm) were tested to determine compliance for creating coronary vessel phantoms. Compliance was assessed using optical coherence tomograph