Unknown

Dataset Information

Accuracy and Interpretation of Transcutaneous Carbon Dioxide Monitoring in Critically Ill Children.


ABSTRACT:

Objectives

Transcutaneous carbon dioxide (Tc co2 ) monitoring can noninvasively assess ventilation by estimating carbon dioxide ( CO2 ) levels in the blood. We aimed to evaluate the accuracy of Tc co2 monitoring in critically ill children by comparing it to the partial pressure of arterial carbon dioxide (Pa co2 ). In addition, we sought to determine the variation between Tc co2 and Pa co2 acceptable to clinicians to modify patient care and to determine which patient-level factors may affect the accuracy of Tc co2 measurements.

Design

Retrospective observational cohort study.

Setting

Single, quaternary care PICU from July 1, 2012, to August 1, 2020.

Patients

Included participants were admitted to the PICU and received noninvasive ventilation support (i.e., cont

SUBMITTER: Setar L 