Ontology highlight
ABSTRACT: Background
Many fentanyl immunoassays are limited in their ability to detect norfentanyl. Urine specimens collected from individuals who have been exposed to fentanyl frequently have detectable concentrations of norfentanyl (≥2 ng/mL) but low concentrations of fentanyl (<2 ng/mL) by LC-MS/MS. The Lin-Zhi Fentanyl II Immunoassay (Lin-Zhi) claims 100% cross-reactivity with norfentanyl and therefore may detect exposure missed by other assays.Methods
In addition to verifying the manufacturer's analytical sensitivity claims, we selected 92 urine specimens with low-positive Lin-Zhi results (1-99 absorbance units, lowest 10%) for analysis by the Immunalysis Health Equity Impact Assessment and ARK II fentanyl methods. The accuracy of the 3 immunoassays was compared to LC-MS/MS as t
SUBMITTER: Uljon S
PROVIDER: S-EPMC11371532 | biostudies-literature | 2024 Sep
REPOSITORIES: biostudies-literature