ABSTRACT: Background
EDTA-induced pseudothrombocytopenia (PTCP) during whole blood collection requires significant laboratory resources to obtain accurate results. We evaluated platelet-deaggregation function in EDTA-induced PTCP and platelet-clump flagging by the BC-6800Plus hematology analyzer using integrated digital image analysis.Methods
We prospectively collected 132 whole blood samples suspected of platelet clumping (102 in EDTA and 30 in sodium citrate) from 88 individuals. We compared platelet counts determined using the platelet count by impedance (PLT-I) function of the DxH 900 hematology analyzer and the PLT-I or optical platelet count (PLT-O) function of the BC-6800Plus. Platelet clumping was verified through manual inspection and the MC-80 digital image analyzer.Res