Combining Enhanced Resolving Power with Duty Cycle Improvements on a Multi-Reflecting Time-of-Flight Mass Spectrometer.
Ontology highlight
ABSTRACT: The combination of enhanced resolving power and improved duty cycle on a multireflecting time-of-flight mass spectrometer is described. Resolving power increases are achieved by extending the effective ion path length from 47 m to greater than 200 m. Path length increases are achieved through containment of ions within the analyzer for up to N = 5 passes using a pulsed deflection electrode. Resolving power was shown to increase from 220,000 to 402,000 (fwhm) at m/z 785 for N = 1 and N = 4 analyzer passes, respectively. Due to the timing of the pulsed deflection electrode, the approach is particularly suited to high resolution analysis over a targeted m/z range. Duty cycle enhancements are achieved for ions of the targeted m/z r