Electric Field-Modulated Electrospray Ionization Mass Spectrometry for Quantity Calibration and Mass Tracking.


ABSTRACT: Analyses conducted by electrospray ionization (ESI) mass spectrometry (MS) typically entail performing a number of preparatory steps, which include quantity calibration and mass calibration. Quantity calibration can be affected by signal noise, while mass calibration can be affected by instrumental drift if analyses are performed over an extended period of time. Here, we present two methods for achieving these calibrations using modulation of electrospray plume by alternating electric fields and demodulating the resulting MS ion currents. For this purpose, we use an ESI source fitted with three ring electrodes between the electrospray emitter and the mass spectrometer's inlet. One of these electrodes is supplied with a sine electric signal. Optionally, a nanoESI emitter is also placed betw