On the extension of the use of a standard operating procedure for nicotine, glycerol and propylene glycol analysis in e-liquids using mass spectrometry.


ABSTRACT:

Introduction

Standard operating procedures (SOP), accessible to several laboratories, are essential for product verification. EU-JATC (European-Joint Action on Tobacco Control) SOP and the WHO TobLabNet (World Health Organization Tobacco Laboratories Network) SOP (SOP11) are available standard methodologies to measure nicotine, glycerol, and propylene glycol, and propose mass spectrometer (MS) as an alternative method to flame ionization detector (FID). This study conducted a comparison between FID and MS concentration results, following the MS method described in SOP11.

Methods

In May 2020, five test e-liquids in replicates (A-E) were prepared at the Istituto di Ricerche Farmacologiche Mario Negri and sent, with SOP 11, validation document and results datasheet to 32 differ