High-Resolution Drift Tube Ion Mobility Spectrometer with Ultra-Fast Polarity Switching.
Ontology highlight
ABSTRACT: Besides safety and security applications, ion mobility spectrometry (IMS) is increasingly used in other fields such as medicine, environmental monitoring and food quality analysis. However, some applications require gas chromatographic separation before analysis by IMS. Furthermore, different compounds in the sample may form positive or negative ions during ionization and therefore simultaneous detection of both ion polarities is highly beneficial to avoid two chromatographic runs of the same sample. This can be achieved by ultra-fast polarity switching of a single drift tube IMS, allowing for quasi-simultaneous detection of both ion polarities. By using a ramped aperture voltage during the switching process, we overcome the issue of excessive displacement currents at the detector during p
SUBMITTER: Hitzemann M
PROVIDER: S-EPMC11391402 | biostudies-literature | 2024 Sep
REPOSITORIES: biostudies-literature