X-Band Single Chip Integrated Pulsed Electron Spin Resonance Microsystem.
Ontology highlight
ABSTRACT: We report on the design and characterization of a single chip integrated pulsed electron spin resonance detector operating at 9.1 GHz. The microsystem consists of an excitation microcoil, a detection microcoil, a low noise microwave preamplifier, a mixer, and an intermediate frequency (IF) amplifier. The chip area is about 0.7 mm2. To exemplify its possible applications, we report the results of single pulse, Rabi nutation, Hahn echo, two echoes, Carr-Purcell, and inversion recovery echo experiments performed on 0.02 and 0.05 nL samples of α, γ-bisdiphenylene-β-phenylallyl (BDPA) and 1% BDPA in polystyrene (BDPA:PS) at room temperature. The measured spin sensitivity is about 8 × 107 spins/Hz1/2 on a sensitive volume of about 0.1 nL. The microsystem power co