Handheld Ultrasound Devices Used by Newly Certified Operators for Pneumonia in the Emergency Department-A Diagnostic Accuracy Study.


ABSTRACT: The diagnostic accuracy of handheld ultrasound (HHUS) devices operated by newly certified operators for pneumonia is unknown. This multicenter diagnostic accuracy study included patients prospectively suspected of pneumonia from February 2021 to February 2022 in four emergency departments. The index test was a 14-zone focused lung ultrasound (FLUS) examination, with consolidation with air bronchograms as diagnostic criteria for pneumonia. FLUS examinations were performed by newly certified operators using HHUS. The reference standard was computed tomography (CT) and expert diagnosis using all medical records. The sensitivity and specificity of FLUS and chest X-ray (CXR) were compared using McNemar's test. Of the 324 scanned patients, 212 (65%) had pneumonia, according to the expert diagnos