Prospective observational study of 2 wearable strain sensors for measuring the respiratory rate.
Ontology highlight
ABSTRACT: The respiratory rate is an important factor for assessing patient status and detecting changes in the severity of illness. Real-time determination of the respiratory rate will enable early responses to changes in the patient condition. Several methods of wearable devices have enabled remote respiratory rate monitoring. However, gaps persist in large-scale validation, patient-specific calibration, standardization and their usefulness in clinical practice has not been fully elucidated. The aim of this study was to evaluate the accuracy of 2 wearable stretch sensors, C-STRECH® which is used in clinical practice and a novel stretchable capacitor in measuring the respiratory rate. The respiratory rate of 20 healthy subjects was measured by a spirometer with the stretch sensor applied to 1 of 5