Large Field-of-View Intravascular Ultrasound for Mitral and Tricuspid Valve-in-Valve Guidance: A Pilot Study.


ABSTRACT:

Background

Actual expansion of a transcatheter heart valve (THV) might differ from nominal, particularly during nonaortic valve-in-valve for a degenerated bioprosthetic surgical heart valve (SHV). This pilot study compared THV expansion measured using large-field-of-view intravascular ultrasound (IVUS) vs. multi-slice computed tomography (MSCT) and assessed the correlation between THV dimensions and transvalvular gradients.

Methods

Fourteen patients were successfully treated with mitral/tricuspid valve-in-valve SAPIEN 3 implantation sized using the true SHV inner diameter; all 14 had baseline MSCT and transvalvular gradients measured at baseline, postprocedure, and at discharge. Periprocedural IVUS (in 6 patients using a Philips 10MHz Vision PV035) was compared with postproc