Combined effect of intermittent hemostasis and a modified external hemorrhage control device in a lethal swine model.


ABSTRACT:

Background

Non-compressible torso hemorrhage (NCTH) presents the ultimate challenge in pre-hospital care. While external hemorrhage control devices (EHCDs) such as the Abdominal Aortic and Junctional Tourniquet (AAJT) and SAM Junctional Tourniquet (SJT) have been invented, the current design and application strategy requires further improvement. Therefore, researchers devised a novel apparatus named Modified EHCD (M-EHCD) and implemented intermittent hemostasis (IH) as a preventive measure against ischemia-reperfusion injury. The objective of this study was to ascertain the combined effect of M-EHCD and IH on the hemostatic effect of NCTH.

Methods

Eighteen swine were randomized to M-EHCD, AAJT or SJT. The NCTH model was established by inducing Class Ⅲ hemorrhagic shock and p

SUBMITTER: Zhang HY 

PROVIDER: S-EPMC11408843 | biostudies-literature | 2024 Sep