Unknown

Dataset Information

Esophageal oxyhemoglobin saturation as a resuscitative metric in hemorrhagic shock.


ABSTRACT:

Background

Mixed venous saturation (SvO2) is considered the gold standard to assess the adequacy of tissue oxygen delivery (DO2) in shock states. However, SvO2 monitoring is challenging as it requires an invasive catheter and frequent blood sampling. Non-invasive methods, including near-infrared spectroscopy, have demonstrated low sensitivity to tissue dysoxia.

Methods

We fabricated a new device that uses resonance Raman spectroscopy (RRS) to quantify oxyhemoglobin saturation (ShbO2) in the esophagus (eShbO2), tongue (tShbO2), and liver (hShbO2). In two rat models of hemorrhagic shock, we quantified (1) The correlation of RRS-measured ShbO2 to SvO2 during progressive hemorrhage (n=20) and (2) The value of these metrics to predict near-term mortality in fixed, severe hemorrha

SUBMITTER: Garcia Mancebo J 

PROVIDER: S-EPMC11409389 | biostudies-literature | 2024