Ontology highlight
ABSTRACT: Background
Mixed venous saturation (SvO2) is considered the gold standard to assess the adequacy of tissue oxygen delivery (DO2) in shock states. However, SvO2 monitoring is challenging as it requires an invasive catheter and frequent blood sampling. Non-invasive methods, including near-infrared spectroscopy, have demonstrated low sensitivity to tissue dysoxia.Methods
We fabricated a new device that uses resonance Raman spectroscopy (RRS) to quantify oxyhemoglobin saturation (ShbO2) in the esophagus (eShbO2), tongue (tShbO2), and liver (hShbO2). In two rat models of hemorrhagic shock, we quantified (1) The correlation of RRS-measured ShbO2 to SvO2 during progressive hemorrhage (n=20) and (2) The value of these metrics to predict near-term mortality in fixed, severe hemorrha
SUBMITTER: Garcia Mancebo J
PROVIDER: S-EPMC11409389 | biostudies-literature | 2024
REPOSITORIES: biostudies-literature