A retrospective characterization of pediatric facemasks marketed in the United States and implications for future designs.


ABSTRACT:

Background

Device manufacturers who seek to market their pediatric facemasks in the United States (U.S.), as part of anthropometric data requirement, need to demonstrate their mask designs are expected to fit the intended user population. However, currently there are no well accepted test methodologies for pediatric facemasks. In addition, unlike N95 respirators, the expected maximum flow rate, and the pressure drop at that expected maximum flow rate for pediatric facemasks have not been established.

Method

The objective of this article is three-fold; use a literature survey to determine the worst-case flow rate, and an acceptable breathing resistance; and come up with a bench-test based protocol for assessing fit of pediatric facemasks.

Results & discussion

The worst