Effects of Loading Forces, Loading Positions, and Splinting of Two, Three, or Four Ti-Zr (Roxolid®) Mini-Implants Supporting the Mandibular Overdentures on Peri-Implant and Posterior Edentulous Area Strains.
Ontology highlight
ABSTRACT: Clinical indications for the Ti-Zr alloy (Roxolid®) mini-implants (MDIs) in subjects with narrow ridges are still under review. The aim was to analyze peri-implant and posterior edentulous area strains dependent on the MDI number, splinting status, loading force, and loading position. Six models were digitally designed and printed. Two, three, or four Ti-Zr MDIs, splinted with a bar or unsplinted (single units), supported mandibular overdentures (ODs), loaded with 50-300 N forces unilaterally, bilaterally, and anteriorly. The artificial mucosa thickness was 2 mm. Strain gauges were bonded on the vestibular and oral peri-implant sides of each MDI, and on the posterior edentulous area under the ODs. Loadings were performed through the metal plate placed on ODs' artificial teeth (1