First in-Lab Testing of a Cost-Effective Prototype for PM<sub>2.5</sub> Monitoring: The P.ALP Assessment.


ABSTRACT: The goal of the present research was to assess, under controlled laboratory conditions, the accuracy and precision of a prototype device (named 'P.ALP': Ph.D. Air-quality Low-cost Project) developed for PM2.5 concentration level monitoring. Indeed, this study follows a complementary manuscript (previously published) focusing on the in-field evaluation of the device's performance. Four P.ALP prototypes were co-located with the reference instrument in a calm-air aerosol chamber at the NIOSH laboratories in Pittsburgh, PA (USA), used by the Center for Direct Reading and Sensor Technologies. The devices were tested for 10 monitoring days under several exposure conditions. To evaluate the performance of the prototypes, different approaches were employed. After the data from the devic