Protein analysis using capillary electrophoresis coupled to mass spectrometry through vibrating sharp-edge spray ionization.
Ontology highlight
ABSTRACT: Capillary electrophoresis (CE) interfaced to mass spectrometry (MS) with electrospray ionization typically incorporates acidic additives or organic solvents to assist in ionization. Vibrating sharp-edge spray ionization (VSSI) is a voltage-free method to interface CE and MS that does not require these additives, making it appealing for protein analyses. CE-VSSI nanoflow sheath separations are performed with low ionic strength aqueous solutions in the sheath to reduce suppression. Serine is also included in the sheath to reduce analyte adduction. Proteins are detected in the 2.5-10 µM range, corresponding to an injected mass range of 0.1-1.2 ng. The anionic proteins β-lactoglobulin and transferrin are resolved using an unmodified fused silica capillary because they do not exhibit nonspecifi