Non-invasive detection of bilirubin concentrations during the first week of life in a low-resource setting along the Thailand-Myanmar border.


ABSTRACT:

Background

Neonatal hyperbilirubinaemia (NH) is a common problem worldwide and is a cause of morbidity and mortality especially in low-resource settings.

Methods

A study was carried out at Shoklo Malaria Research Unit (SMRU) clinics along the Thailand-Myanmar border to evaluate a non-invasive test for diagnosis of NH in a low-resource setting. Performance of a transcutaneous bilirubinometer Dräger Jaundice Meter JM-105 was assessed against routine capillary serum bilirubin testing (with BR-501 microbilirubinometer) before phototherapy during neonatal care in the first week of life. Results were analysed by direct agreement and by various bilirubin thresholds used in clinical practice. Total serum bilirubin was also measured in cord blood at birth and tested for prediction of