Biocompatibility analysis of titanium bone wedges coated by antibacterial ceramic-polymer layer.
Ontology highlight
ABSTRACT: This paper presents the surface treatment results of titanium, veterinary bone wedges. The functional coating is composed of a porous oxide layer (formed by a plasma electrolytic oxidation process) and a polymer poly(sebacic anhydride) (PSBA) layer loaded with amoxicillin (formed by dip coatings). The coatings were porous and composed of Ca (4.16%-6.54%) and P (7.64%-9.89% determined by scanning electron microscopy with EDX) in the upper part of the implant. The titanium bone wedges were hydrophilic (54° water contact angle) and rough (surface area (Sa):1.16 μm) The surface tension determined using diiodomethane was 68.6 ± 2.0° for the anodized implant and was similar for hybrid coatings: 60.7 ± 2.2°. 12.87 ± 0.91 µg/mL of amoxicillin was released from the implants during the first 30 min
SUBMITTER: Kazek-Kesik A
PROVIDER: S-EPMC11452723 | biostudies-literature | 2024 Oct