Unknown

Dataset Information

Breaking through diabetes management barriers: Assessing the accuracy of Freestyle Libre Pro flash continuous glucose monitoring in diabetes patients with myocardial infarction.


ABSTRACT:

Purpose

The objective of this study was to assess the accuracy of FreeStyle Libre Pro (FSL-Pro) flash continuous glucose monitoring (CGM) in patients with type 2 diabetes mellitus (T2DM) and acute myocardial infarction (AMI).

Methods

A single-arm, single-center prospective study was conducted in the cardiac care unit from January 2021 to September 2023. Patients underwent finger-prick blood glucose (FPBG) testing before breakfast (6:00 am) and after meals (at 9:00, 13:00, 19:00 pm), along with CGM during their hospitalization. Statistical analyses included mean differences (MDs), mean absolute relative difference (MARDs) of blood glucose levels, and hypoglycemia occurrences. A Bland-Altman plot analysis and Pearson correlation were performed.

Results

Ninety-seven T2DM

SUBMITTER: Shen L 

PROVIDER: S-EPMC11456750 | biostudies-literature | 2024 Jan-Dec

REPOSITORIES: biostudies-literature