ABSTRACT: Background
Identifying a weak/strong pelvic floor muscle (PFM) contraction in pregnant women may help prevent and treat dysfunctions during late pregnancy and postpartum.Objective
To determine whether the Peritron™ manometer can accurately differentiate a weak from a strong PFM contraction and the respective cut-offs for its variables in pregnant women.Methods
This is a diagnostic accuracy study. Forty-four women in the third trimester of pregnancy participated (mean±SD age: 29±5 years). The reference test was vaginal palpation, and the index test was vaginal manometry (Peritron™ manometer). Variables assessed by vaginal manometry were rest, maximal voluntary contraction (MVC), MVC average, duration, gradient, area under the curve (AUCm), and contraction speed. The