Biomechanical Analysis of the Ross Procedure in an Ex Vivo Left Heart Simulator.


ABSTRACT:

Background

Neo-aortic pulmonary autografts often experience root dilation and valve regurgitation over time. This study seeks to understand the biomechanical differences between aortic and neo-aortic pulmonary roots using a heart simulator.

Methods

Porcine aortic, neo-aortic pulmonary, and pulmonary roots (n  =  6) were mounted in a heart simulator (parameters: 100 mm Hg, 37 °C, 70 cycles per minute, 5.0 L/min cardiac output). Echocardiography was used to study root distensibility (percentage change in luminal diameter between systole and diastole) and valve function. Leaflet motion was tracked with high-speed videography. After 30 min in the simulator, leaflet thickness (via cryosectioning), and multiaxial modulus (via lenticular hydrostatic deformation testing) were obtain