ABSTRACT: Purpose
To assess the impact of working distance (WD) on optical distortion in optical coherence tomography (OCT) imaging and to evaluate the effectiveness of optical distortion correction in achieving consistent retinal Bruch's membrane (BM) layer curvature, regardless of variations in WD.Methods
Ten subjects underwent OCT imaging with four serial macular volume scans, each employing distinct WD settings adjusted by balancing the sample and reference arm of the OCT interferometer (eye length settings changed). Either of two types of 30° standard objectives (SOs) was used. A ray tracing model was used to correct optical distortion, and BM layer curvature (represented as the second derivative of the curve) was measured. Linear mixed effects (LME) modeling was used to analyze