Unknown

Dataset Information

Validation and correlation of high-sensitive troponin I and troponin T in the emergency department.


ABSTRACT:

Background

Troponin elevation is frequently observed in various scenarios in the Emergency Department (ED), yet there is a paucity of studies investigating simultaneously measured high-sensitivity cardiac troponin T (hs-cTnT) and troponin I (hs-cTnI) within a diverse cohort in a clinical setting.

Methods

All patients who underwent troponin testing at a single center were eligible for this study. Only patients with simultaneous samples with hs-cTnI (Siemens) and hs-cTnT (Roche) were included, regardless of chief complaint.

Results

Analysis of 1987 samples from 1134 patients showed a significant correlation between hs-cTnT and hs-cTnI (r = 0.86, p < 0.01). Of these samples, 65% exceeded the upper reference limit (URL) for hs-cTnT, and 30% for hs-cTnI with 39% who exhibi

SUBMITTER: Nseir M 

PROVIDER: S-EPMC11470706 | biostudies-literature | 2024 Oct

REPOSITORIES: biostudies-literature