Comparison between a Flash Glucose Monitoring System and a Portable Blood Glucose Meter for Monitoring of Cats with Diabetic Ketosis or Ketoacidosis.


ABSTRACT: Cats with diabetic ketosis or ketoacidosis DK(A) require intensive glucose monitoring. The aim of this study was to assess the agreement between a portable blood glucose meter (PBGM) and a flash glucose monitoring system (FGMS; FreeStyle Libre 2.0 Abbott®) measuring interstitial glucose in cats with DK(A). Ten client-owned cats with naturally occurring DK(A) were prospectively enrolled. Glucose concentrations were assessed with both methods every 1-3 h until resolution of DK(A), and every 4-8 h thereafter. While the median FGMS measured glucose concentration (14.3 mmol/L) was significantly lower than the median PBGM measured glucose concentration (19 mmol/L) (p < 0.001), the overall correlation between the FGMS and PBGM was high (r = 0.88; p < 0.001). In the Parkes